Premium
Determination of ammonia, creatinine and inorganic cations in urine using CE with contactless conductivity detection
Author(s) -
Tůma Petr,
Samcová Eva,
Duška František
Publication year - 2008
Publication title -
journal of separation science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.72
H-Index - 102
eISSN - 1615-9314
pISSN - 1615-9306
DOI - 10.1002/jssc.200700655
Subject(s) - chemistry , chromatography , creatinine , acetic acid , urine , ammonia , analyte , analytical chemistry (journal) , biochemistry
Abstract CE with capacitively coupled contactless conductivity detection (C 4 D) was used to determine waste products of the nitrogen metabolism (ammonia and creatinine) and of biogenic inorganic cations in samples of human urine. The CE separation was performed in two BGEs, consisting of 2 M acetic acid + 1.5 mM crown ether 18‐crown‐6 (BGE I) and 2 M acetic acid + 2% w/v PEG (BGE II). Only BGE II permitted complete separation of all the analytes in a model sample and in real urine samples. The LOD values for the optimized procedure ranged from 0.8 μM for Ca 2+ and Mg 2+ to 2.9 μM for NH 4 + (in terms of mass concentration units, from 7 μg/L for Li + to 102 μg/L for creatinine). These values are adequate for determination of NH 4 + , creatinine, Na + , K + , Ca 2+ and Mg 2+ in real urine samples.