Premium
Warping of cross sections in the torsion of long bones with internal fracture fixation plates
Author(s) -
Beaupré G. S.,
Carter D. R.
Publication year - 1987
Publication title -
journal of orthopaedic research
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.041
H-Index - 155
eISSN - 1554-527X
pISSN - 0736-0266
DOI - 10.1002/jor.1100050216
Subject(s) - image warping , torsion (gastropod) , internal fixation , orthodontics , anatomy , geology , computer science , artificial intelligence , medicine
Torsion of noncircular beams results in warping of each cross section. When noncircular cross sections are constrained to remain plane, the resulting shear stress distribution is different from what Saint Venant torsion (with warping) would predict. This has practical implications to the stress analysis of plated long bones subjected to torsional loadings. Analyses in which warping is not allowed predict incorrect stress fields in the plate and bone and overpredict the amount of stress shielding associated with fracture plate fixation.