Premium
Electrical osteogenesis by low direct current
Author(s) -
Baranowski Thomas J.,
Black Jonathan,
Brighton Carl T.,
Friedenberg Zachary B.
Publication year - 1983
Publication title -
journal of orthopaedic research
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.041
H-Index - 155
eISSN - 1554-527X
pISSN - 0736-0266
DOI - 10.1002/jor.1100010202
Subject(s) - cathode , medullary cavity , electrode , current (fluid) , tibia , direct current , chemistry , materials science , electric current , biomedical engineering , anatomy , electrical engineering , medicine , voltage , engineering
A constant direct current cathode was employed in the medullary canal of the rabbit tibia to investigate electrical osteogenesis at low current levels. Currents of 0.015 or 0.075 μA were delivered to the bone and the biological response was compared with contralateral controls receiving 20 μA. This investigation was performed to determine if electrical osteogenesis occurs at current levels below the previously studied range of 1–100 μA with stainless steel electrodes. New bone formed by 0.015 μA cathodes was statistically comparable with that found around inactive cathodes from an earlier pertinent study. The osteogenic response to 0.075 μA cathodes was significantly elevated above that to inactive ones, thus substantiating electrical osteogenesis for currents below 1 μA. However, it is evident that this does not demonstrate a further stimulatory range but that currents near 0.075 μA probably approach the lower significant limit for electrically induced bone growth with stainless steel electrodes.