Comparison of SARS‐CoV‐2 detection from nasopharyngeal swab samples by the Roche cobas 6800 SARS‐CoV‐2 test and a laboratory‐developed real‐time RT‐PCR test
Author(s) -
Pujadas Elisabet,
Ibeh Nnaemeka,
Hernandez Matthew M.,
Waluszko Aneta,
Sidorenko Tatyana,
Flores Vanessa,
Shiffrin Biana,
Chiu Numthip,
YoungFrancois Alicia,
Nowak Michael D.,
PanizMondolfi Alberto E.,
Sordillo Emilia M.,
CordonCardo Carlos,
Houldsworth Jane,
Gitman Melissa R.
Publication year - 2020
Publication title -
journal of medical virology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.782
H-Index - 121
eISSN - 1096-9071
pISSN - 0146-6615
DOI - 10.1002/jmv.25988
Subject(s) - covid-19 , virology , concordance , medicine , real time polymerase chain reaction , sars virus , false negative reactions , confidence interval , biology , disease , infectious disease (medical specialty) , biochemistry , gene
The urgent need to implement and rapidly expand testing for severe acute respiratory syndrome coronavirus‐2 (SARS‐CoV‐2) infection has led to the development of multiple assays. How these tests perform relative to one another is poorly understood. We evaluated the concordance between the Roche Diagnostics cobas 6800 SARS‐CoV‐2 test and a laboratory‐developed test (LDT) real‐time reverse transcription‐polymerase chain reaction based on a modified Centers for Disease Control and Prevention protocol, for the detection of SARS‐CoV‐2 in samples submitted to the Clinical Laboratories of the Mount Sinai Health System. A total of 1006 nasopharyngeal swabs in universal transport medium from persons under investigation were tested for SARS‐CoV‐2 as part of routine clinical care using the cobas SARS‐CoV‐2 test with subsequent evaluation by the LDT. Cycle threshold values were analyzed and interpreted as either positive (“detected” or “presumptive positive”), negative (not detected), inconclusive, or invalid. Statistical analysis was performed using GraphPad Prism 8. The cobas SARS‐CoV‐2 test reported 706 positive and 300 negative results. The LDT reported 640 positive, 323 negative, 34 inconclusive, and 9 invalid results. When excluding inconclusive and invalid results, the overall percent agreement between the two platforms was 95.8%. Cohen's κ coefficient was 0.904 (95% confidence interval, 0.875‐0.933), suggesting almost perfect agreement between both platforms. An overall discordance rate of 4.2% between the two systems may reflect differences in primer sequences, assay limit of detection, or other factors, highlighting the importance of comparing the performance of different testing platforms.