Formation of solvated ions in the atmospheric interface of an electrospray ionization triple‐quadrupole mass spectrometer
Author(s) -
Schlosser Gitta,
Takáts Zoltán,
Vékey Károly
Publication year - 2003
Publication title -
journal of mass spectrometry
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.475
H-Index - 121
eISSN - 1096-9888
pISSN - 1076-5174
DOI - 10.1002/jms.555
Subject(s) - chemistry , solvation , quadrupole mass analyzer , mass spectrometry , acetonitrile , ion , solvent , electrospray ionization , alkali metal , triple quadrupole mass spectrometer , molecule , ionization , quadrupole , ion trap , inorganic chemistry , analytical chemistry (journal) , tandem mass spectrometry , selected reaction monitoring , organic chemistry , chromatography , atomic physics , physics
Abstract A simple method capable of generating and investigating various solvent clusters and solvated ions was developed. The technique opens a door to studying these complexes on commercially available instruments. Formation of the desired solvated ion in the gas phase was achieved by introducing the appropriate volatile solvent vapour into the curtain gas stream. Capabilities of the technique are illustrated by generating alkali, alkaline earth and transition metal cations solvated by various volatile compounds such as water, methanol and acetonitrile. Depending on the ligands and on the experimental conditions, clusters of 2–100 molecules may be observed. Isotope labelling suggests that these are formed by a re‐solvation process in the curtain gas region. Copyright © 2003 John Wiley & Sons, Ltd.