z-logo
Premium
Comparison of technical failure of MR elastography for measuring liver stiffness between gradient‐recalled echo and spin‐echo echo‐planar imaging: A systematic review and meta‐analysis
Author(s) -
Kim Dong Wook,
Kim So Yeon,
Yoon Hee Mang,
Kim Kyung Won,
Byun Jae Ho
Publication year - 2020
Publication title -
journal of magnetic resonance imaging
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.563
H-Index - 160
eISSN - 1522-2586
pISSN - 1053-1807
DOI - 10.1002/jmri.26918
Subject(s) - magnetic resonance elastography , meta analysis , confidence interval , medicine , subgroup analysis , elastography , nuclear medicine , radiology , ultrasound
Background Magnetic resonance elastography (MRE) using a gradient‐recalled echo (GRE) or a recently available spin‐echo echo‐planar imaging (SE‐EPI) sequence is a promising noninvasive method for measuring liver stiffness. However, it sometimes fails to measure stiffness values, thereby resulting in technical failures. Purpose To assess and compare technical failures of MRE for measuring liver stiffness between GRE and SE‐EPI sequences. Study Type Systematic review and meta‐analysis. Population Eight studies with both GRE and SE‐EPI, 22 studies with only GRE, one study with only SE‐EPI. Field Strength/Sequence Either 1.5 or 3T MRE using GRE and/or SE‐EPI. Assessment Through an Ovid‐MEDLINE and EMBASE database search, original articles investigating the proportion of MRE technical failures in the measurement of liver stiffness published up until October 2018 were screened and selected. Statistical Analysis The pooled proportions of technical failures under GRE and SE‐EPI were calculated using random‐effects meta‐analysis of single proportions and inverse variance for calculating weights. Subgroup analyses were performed to explore the covariates affecting heterogeneity. Head‐to‐head comparisons of technical failure between the sequences were conducted with eight MRE studies using both GRE and SE‐EPI. Results The pooled proportion of technical failure under GRE MRE was 5.8% (95% confidence interval [CI], 4.6–7.4%), and a subgroup analysis showed higher technical failure rates at 3T than at 1.5T. The pooled proportion of technical failure under SE‐EPI MRE was 2.0% (95% CI, 1.3–3.4%), without significant differences ( P = 0.38–0.89) being observed in the subgroup analyses. In the eight studies comparing the two sequences, failure was more frequently observed with GRE than with SE‐EPI (9.4% vs. 1.9%; P  < 0.01). Data Conclusion MRE conducted with SE‐EPI sequences showed a lower technical failure rate than GRE sequences. With GRE sequences, a magnetic field of 3T was associated with higher technical failure rates than was 1.5T. Level of Evidence: 1 Technical Efficacy Stage: 3 J. Magn. Reson. Imaging 2020;51:1086–1102.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here