In vitro assessment of knee MRI in the presence of metal implants comparing MAVRIC‐SL and conventional fast spin echo sequences at 1.5 and 3 T field strength
Author(s) -
Liebl Hans,
Heilmeier Ursula,
Lee Sonia,
Nardo Lorenzo,
Patsch Janina,
Schuppert Christopher,
Han Misung,
Rondak InaChristine,
Banerjee Suchandrima,
Koch Kevin,
Link Thomas M.,
Krug Roland
Publication year - 2015
Publication title -
journal of magnetic resonance imaging
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.563
H-Index - 160
eISSN - 1522-2586
pISSN - 1053-1807
DOI - 10.1002/jmri.24668
Subject(s) - fast spin echo , magnetic resonance imaging , nuclear medicine , biomedical engineering , materials science , artifact (error) , medicine , radiology , computer science , computer vision
Purpose To assess lesion detection and artifact size reduction of a multiacquisition variable‐resonance image combination, slice encoding for metal artifact correction (MAVRIC‐SEMAC) hybrid sequence (MAVRIC‐SL) compared to standard sequences at 1.5T and 3T in porcine knee specimens with metal hardware. Materials and Methods Artificial cartilage and bone lesions of defined size were created in the proximity of titanium and steel screws with 2.5 mm diameter in 12 porcine knee specimens and were imaged at 1.5T and 3T magnetic resonance imaging (MRI) with MAVRIC‐SL PD and short T 1 inversion recovery (STIR), standard fast spin echo (FSE) T 2 PD, and STIR and fat‐saturated T 2 FSE sequences. Three radiologists blinded to the lesion locations assessed lesion detection rates on randomized images for each sequence using receiver operating characteristic (ROC). Artifact length and width were measured. Results Metal artifact sizes were largest in the presence of steel screws at 3T (FSE T 2 FS: 28.7 cm 2 ) and 1.5T (16.03 cm 2 ). MAVRIC‐SL PD and STIR reduced artifact sizes at both 3T (1.43 cm 2 ; 2.46 cm 2 ) and 1.5T (1.16 cm 2 ; 1.59 cm 2 ) compared to FS T 2 FSE sequences (27.57 cm 2 ; 13.20 cm 2 ). At 3T, ROC‐derived AUC values using MAVRIC‐SL sequences were significantly higher compared to standard sequences (MAVRIC‐PD: 0.87, versus FSE‐ T 2 ‐FS: 0.73 [ P = 0.025]; MAVRIC‐STIR: 0.9 vs. T 2 ‐STIR: 0.78 [ P = 0.001] and vs. FSE‐ T 2 ‐FS: 0.73 [ P = 0.026]). Similar values were observed at 1.5T. Comparison of 3T and 1.5T showed no significant differences (MAVRIC‐SL PD: P = 0.382; MAVRIC‐SL STIR: P = 0.071). Conclusion MAVRIC‐SL sequences provided superior lesion detection and reduced metal artifact size at both 1.5T and 3T compared to conventionally used FSE sequences. No significant disadvantage was found comparing MAVRIC‐SL at 3T and 1.5T, although metal artifacts at 3T were larger. J. Magn. Reson. Imaging 2015;41:1291–1299. © 2014 Wiley Periodicals, Inc.