Correction of frequency drifts induced by gradient heating in 1H spectra using interleaved reference spectroscopy
Author(s) -
Lange Thomas,
Zaitsev Maxim,
Buechert Martin
Publication year - 2011
Publication title -
journal of magnetic resonance imaging
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.563
H-Index - 160
eISSN - 1522-2586
pISSN - 1053-1807
DOI - 10.1002/jmri.22471
Subject(s) - spectroscopy , nuclear magnetic resonance , spectral line , analytical chemistry (journal) , materials science , physics , chemistry , chromatography , quantum mechanics , astronomy
Abstract Purpose: To evaluate the impact of heating‐induced frequency drifts on single‐voxel spectroscopy and to demonstrate correction strategies based on the interleaved reference scan technique. Materials and Methods: Frequency drifts induced by gradient heating are assessed for two clinical 3 Tesla (T) whole body MR systems. The interleaved reference scan (IRS) method is used for correcting these frequency drifts in 1H spectra in vitro and in vivo. For severely drift‐affected spectroscopy experiments, a feedback‐based version of the IRS sequence is proposed, which adds the functionality of a frequency lock to prevent a degradation of the water suppression. Results: It is shown that the line widths of the spectral resonances can be largely reduced with the interleaved reference scan method, resulting in considerably improved peak resolution and signal‐to‐noise ratio. The feedback‐based IRS method additionally allows for stable water suppression, even in the presence of very strong frequency drifts. Conclusion: If spectroscopy scans are combined with imaging scans with a high gradient duty cycle such as diffusion‐weighted imaging or functional MRI, a drift correction with IRS can considerably improve the validity of data analysis in research studies. J. Magn. Reson. Imaging 2011;33:748–754. © 2011 Wiley‐Liss, Inc.