Impact of motion on T1 mapping acquired with inversion recovery fast spin echo and rapid spoiled gradient recalled‐echo pulse sequences for delayed gadolinium‐enhanced MRI of cartilage (dGEMRIC) in volunteers
Author(s) -
Studler Ueli,
White Lawrence M.,
Andreisek Gustav,
Luu Sheena,
Cheng HaiLing Margaret,
Sussman Marshall S.
Publication year - 2010
Publication title -
journal of magnetic resonance imaging
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.563
H-Index - 160
eISSN - 1522-2586
pISSN - 1053-1807
DOI - 10.1002/jmri.22249
Subject(s) - fast spin echo , magnetic resonance imaging , nuclear medicine , spin echo , nuclear magnetic resonance , medicine , flip angle , echo (communications protocol) , physics , radiology , computer science , computer network
Purpose: To evaluate the impact of motion on T1 values acquired by using either inversion‐recovery fast spin echo (IR‐FSE) or three‐dimensional (3D) spoiled gradient recalled‐echo (SPGR) sequences for delayed gadolinium‐enhanced magnetic resonance imaging of cartilage (dGEMRIC) in volunteers. Materials and Methods: Single‐slice IR‐FSE and 3D SPGR sequences were applied to perform dGEMRIC in five healthy volunteers. A mutual information‐based approach was used to correct for image misregistration. Displacements were expressed as averaged Euclidean distances and angles. Averages of differences in goodness of fit (Δχ 2 ) tests and averages of relative differences in T1 values (ΔT1) before and after motion correction were computed. Results: Maximum Euclidean distance was 3.5 mm and 1.2 mm for IR‐FSE and SPGR respectively. Mean ± SD of Δχ 2 were 10.18 ± 8.4 for IR‐FSE and −1.37 ± 5.5 for SPGR. Mean ± SD of ΔT1 were 0.008 ± 0.0048 for IR‐FSE and −0.002 ± 0.019 for FSPGR. Pairwise comparison of Δχ 2 values showed a significant difference for IR‐FSE, but not for 3D‐SPGR. Significantly greater variability in T1 values was also noted for IR‐FSE than for 3D‐SPGR. Conclusion: Involuntary motion has a significant influence on T1 values acquired with IR‐FSE, but not with 3D‐SPGR in healthy volunteers. J. Magn. Reson. Imaging 2010;32:394–398. © 2010 Wiley‐Liss, Inc.