Premium
Temperature monitoring during MR imaging: Comparison of fluoroptic and standard thermistors
Author(s) -
Taber Katherine H.,
Hayman L. Anne
Publication year - 1992
Publication title -
journal of magnetic resonance imaging
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.563
H-Index - 160
eISSN - 1522-2586
pISSN - 1053-1807
DOI - 10.1002/jmri.1880020119
Subject(s) - thermistor , materials science , nuclear medicine , medicine , biomedical engineering , electrical engineering , engineering
Temperature readings from standard wire lead thermistor probes and a fluoroptic temperature probe were compared before, during, and after exposure to a spin‐echo imaging sequence at 2.35 T. All the probes were placed in saline‐filled vials (1‐mL capacity) surrounded by dense foam rubber to isolate and preserve local heating effects. There were no consistent differences in the temperatures recorded by the various probes. The authors conclude that wire lead thermistor probes do not exhibit substantial heat deposition when exposed to a standard spin‐echo imaging sequence at 2.35 T.