Evaluation of a novel point‐of‐care test kit, ABSOGEN ™ PCT, in semi‐quantitative measurement of procalcitonin in whole blood
Author(s) -
Kwon Hi Jeong,
Lee Jehoon,
Park Haeil,
Han Kyungja
Publication year - 2017
Publication title -
journal of clinical laboratory analysis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.536
H-Index - 50
eISSN - 1098-2825
pISSN - 0887-8013
DOI - 10.1002/jcla.22111
Subject(s) - procalcitonin , concordance , medicine , point of care testing , cartridge , point of care , whole blood , naked eye , chromatography , chemistry , pathology , detection limit , materials science , metallurgy , sepsis
Background In response to inflammation, procalcitonin plasma concentrations increase more rapidly than other acute‐phase reactants and higher values are associated with severe disease. Procalcitonin measurements assist in determining whether antibiotic therapy should be used. point‐of‐care testing ( POCT ) is performed for early decision making about additional testing or therapy. The ABSOGEN ™ PCT (Bumyoungbio, Inc., Suwon, Korea) is a rapid novel semi‐quantitative immunochromatographic PCT assay that analyses whole blood samples. We compared the patient quantitative test results to ABSOGEN ™ PCT test results. Methods Whole blood was loaded onto an ABSOGEN ™ PCT cartridge and incubated for 10 minutes. The color intensity of the band of the cartridge was measured using an accompanying device and with the naked eye. The results were graded as negative (<0.1 ng/mL), low (0.1 to <1.0 ng/mL), middle (1.0‐2.0 ng/mL), and high (>2.0 ng/mL). A total of 158 specimens with procalcitonin levels measured from 0 to 18.96 ng/mL were used for comparison study. Results The concordance rate between ABSOGEN ™ PCT using the reader and quantitative assay, between ABSOGEN ™ PCT using naked eyes and quantitative assay, and between ABSOGEN ™ PCT using the reader and naked eyes for the same category was 83.5% ( P =.040), 78.5% ( P <.001), and 82.3% ( P =.001), respectively. The concordance rates for the ±1 categories were all 100%. Conclusion ABSOGEN ™ PCT is an accurate assay. It is easy to use, simple, fast, portable. The assay has potential value in clinical applications, especially in emergency care.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here