z-logo
Premium
High‐resolution pluronic‐filled microchip CE‐SSCP analysis system via channel width control
Author(s) -
Shin Giyoung,
Kim DongKyun,
Doh Junsang,
Lee Daeyeon,
Lee Nam Ki,
Jung Gyoo Yeol
Publication year - 2016
Publication title -
electrophoresis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.666
H-Index - 158
eISSN - 1522-2683
pISSN - 0173-0835
DOI - 10.1002/elps.201500427
Subject(s) - resolution (logic) , poloxamer , materials science , channel (broadcasting) , chromatography , analytical chemistry (journal) , chemistry , computer science , composite material , polymer , telecommunications , artificial intelligence , copolymer
Although the resolution of CE‐SSCP has been significantly improved by using a poly(ethyleneoxide)‐poly(propyleneoxide)‐poly(ethyleneoxide) (PEO‐PPO‐PEO; Pluronic ® ) triblock copolymer as a separation medium, CE‐SSCP on a microchip format is not widely applicable because their resolution is limited by short channel length. Therefore, a strategy to improve the resolution in channels of limited lengths is highly required for enabling microchip‐based CE‐SSCP. In this study, we developed a high‐resolution CE‐SSCP microchip system by controlling the width of the pluronic‐filled channel. We tested four different channel widths of 180, 240, 300, and 400 μm, and found that 300 μm showed the highest resolution in the separation of two pathogen specific markers. Potential applications of our method in various genetic analyses were also shown by using SNP markers for spinal muscular atrophy.