Development of a quality control method for the characterization of oligonucleotides by capillary zone electrophoresis‐electrospray ionization‐quadrupole time of flight‐mass spectrometry
Author(s) -
Willems An V.,
Deforce Dieter L.,
Van Peteghem Carlos H.,
Van Bocxlaer Jan F.
Publication year - 2005
Publication title -
electrophoresis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.666
H-Index - 158
eISSN - 1522-2683
pISSN - 0173-0835
DOI - 10.1002/elps.200406201
Subject(s) - electrospray ionization , capillary electrophoresis , chemistry , adduct , chromatography , analytical chemistry (journal) , mass spectrometry , oligonucleotide , electrospray , ammonium acetate , ionization , ion , high performance liquid chromatography , organic chemistry , dna , biochemistry
A capillary zone electrophoresis‐negative electrospray ionization‐quadrupole time of flight‐mass spectrometric method was developed for the characterization of oligonucleotides after synthesis, using model compounds. The major difficulty is the adduction of metal cations to the polyanionic backbone of the oligonucleotide sample, resulting in complex spectra and decreased sensitivity. Several approaches were investigated to circumvent this problem. Separation was performed in an ammonium carbonate buffer. During separation, the interfering metal ions were exchanged for ammonium ions, which are less tightly bound to the oligonucleotide when ionized. The influence of the addition of piperidine and imidazole or trans ‐1,2‐diaminocyclohexane‐ N , N , N' , N' ‐tetraacetic acid (CDTA) to the running buffer for further reduction of cation adduction was investigated. Addition of CDTA to the buffer system resulted in a deconvoluted spectrum with very little adducts. On‐line sample stacking proved vital to preconcentrate the samples. The pH and the concentration of the ammonium carbonate buffer as well as the electrophoresis voltage were optimized to achieve the best signal response for the oligonucleotides and a maximum reduction of the cation adducts as well as a short analysis time. Finally, the sheath liquid composition was examined for further improvement of the signal. The developed method was used to analyze different oligonucleotides (5000–9200 Da) in light of its use as a final quality control method for oligonucleotides in terms of purity and sequence homogeneity of the synthesized products. In all cases, very little adducts were observed in the deconvoluted spectra, and the relative errors of the measured molecular masses ranged from 3 to 35 ppm.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here