z-logo
Premium
Determination of oxalate in human serum by multicolumn isotachophoresis
Author(s) -
Dolník Vladislav,
Deml Mirko,
Boček Petr
Publication year - 1988
Publication title -
electrophoresis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.666
H-Index - 158
eISSN - 1522-2683
pISSN - 0173-0835
DOI - 10.1002/elps.1150091209
Subject(s) - isotachophoresis , chromatography , oxalate , chemistry , electrolyte , suspension (topology) , reproducibility , detection limit , inorganic chemistry , electrode , mathematics , homotopy , pure mathematics
The practical usefulness of multicolumn isotachophoresis was demonstrated by the determination of oxalate levels in human serum. 10 mmol/L HC1 and 100 mmol/L Na 2 HPO 4 served as the leading and terminating electrolyte, respectively. For the stabilization of the isotachophoretic zones in large cross section channels a suspension of Bio‐Gel P‐300 in the leading electrolyte was used. For the analysis ca . 1 mL of fresh serum was required and 1.2 mL of 1:1 mixture of serum and the suspension of Bio‐Gel P‐300 in deionized water was applied as sample. The time of analysis ranged between 45–49 min. The detection limit of analysis was determined to be 50 nmol/L. Reproducibility of analysis of 1 μmol/L oxalate in water standard was found to be 2.6% ( n = 6).