z-logo
Premium
Disposable Biosensor Based on Amidase/CeO2/GNR Modified Inkjet‐printed CNT Electrodes‐droplet Based Paracetamol Detection in Biological Fluids for “Point‐of‐care” Applications
Author(s) -
Stanković Dalibor M.,
Ognjanović Miloš,
Jović Milica,
Cuplić Valentina,
Lesch Andreas,
Girault Hubert H.,
Gavrović Jankulović Marija,
Antić Bratislav
Publication year - 2019
Publication title -
electroanalysis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.574
H-Index - 128
eISSN - 1521-4109
pISSN - 1040-0397
DOI - 10.1002/elan.201900129
Subject(s) - biosensor , detection limit , materials science , electrode , reproducibility , nanotechnology , linear range , graphene , cerium , chromatography , chemistry , metallurgy
A disposable acetaminophen biosensor based on inkjet‐printed CNT electrodes (IJPCNT) modified with amidase/cerium dioxide@graphene nanoribbons composite was developed (ACeO2@GNR/IJPCNT). The enzyme amidase A was used for the first time as a recognition element. Inkjet‐printed CNT electrodes served as a basis for the construction of a biosensor that enables droplet detection using 5 μL sample volume. The biosensor showed high selectivity, sensitivity, a low detection limit of 0.18 μM and a wide working linear range from 1 to 100 μM. The proposed approach allows fast and reliable detection of acetaminophen in biological fluids with negligible matrix effect and remarkable reproducibility.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here