Premium
Amperometric Glucose Biosensors Based on Glassy Carbon and SWCNT‐Modified Glassy Carbon Electrodes
Author(s) -
Carpani Irene,
Scavetta Erika,
Tonelli Domenica
Publication year - 2008
Publication title -
electroanalysis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.574
H-Index - 128
eISSN - 1521-4109
pISSN - 1040-0397
DOI - 10.1002/elan.200704054
Subject(s) - amperometry , biosensor , glucose oxidase , glassy carbon , carbon nanotube , electrode , electrochemistry , carbon fibers , materials science , chemistry , nanotechnology , chemical engineering , cyclic voltammetry , composite material , composite number , engineering
Different carbonaceous materials, such as single‐walled carbon nanotubes (SWCNTs) and glassy carbon submitted to an electrochemical activation at +1.80 V (vs. SCE) for 900 s, have been used with the aim of comparing their performances in the development of enzyme electrodes. Commercial SWCNTs have been pretreated with 2.2 M HNO 3 for 20 h prior to use. The utility of activated GC as promising material for amperometric oxidase‐based biosensors has been confirmed. With glucose oxidase (GOx) as a model enzyme, glucose was efficiently detected up to 1 mM without the use of a mediator. Both electrodes operated in stirred solutions of 0.1 M phosphate buffer (pH 5.5), containing dissolved oxygen, at a potential of −0.40 V vs. SCE. Although the performances of the two carbonaceous materials were comparable, the biosensors based on activated GC were characterized by a practically unchanged response 40 days after the fabrication, a better signal to noise ratio, and a little worse sensitivity. In addition, the preparation procedure of such biosensors was more simple, rapid and reproducible.