Devices external pulse generator: A reliable temporary pacemaker?
Author(s) -
Inoue D.,
Katsume H.,
Ilichi H.,
Ochiai M.
Publication year - 1987
Publication title -
clinical cardiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.263
H-Index - 72
eISSN - 1932-8737
pISSN - 0160-9289
DOI - 10.1002/clc.4960101208
Subject(s) - medicine , pulse generator , generator (circuit theory) , pulse (music) , electrical engineering , voltage , power (physics) , engineering , physics , quantum mechanics
Two cases of malfunctioning APC (American Pacemaker Company) Devices pacemakers are reported. Two patients with sick sinus syndrome were temporarily paced using APC Devices pacemaker models EC 4542 and EV 4543 respectively, which showed transient increase (53% and 83% of the preset rate, respectively) in pacing rate. The sudden increase of the pacing rate might be the pacemaker's design to switch to fixed‐rate pacing at nominally 25% higher than the selected rate in the presence of an excessive level of electrical interference. However, the increased rate was much faster than the interference rate. Great caution should be paid when APC Devices pacemaker is to be used.