Fluctuations of the atrial cycle length and A‐V delay with a universal (DDD) pacemaker
Author(s) -
Bognolo D.,
Vijayanagar R.,
Eckstein P.,
Jeffery D.,
Rosenfield L.
Publication year - 1984
Publication title -
clinical cardiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.263
H-Index - 72
eISSN - 1932-8737
pISSN - 0160-9289
DOI - 10.1002/clc.4960070707
Subject(s) - medicine , cardiology , interval (graph theory) , cardiac cycle , pulse generator , atrioventricular block , voltage , combinatorics , electrical engineering , mathematics , engineering
The electronic characteristics of a new universal (DDD) pulse generator can result in significant shortening of the atrial cycle length (A‐A interval) and shortening of the atrioventricular (A‐V) interval beyond the programmed values. These pacemaker arrhythmias are entirely compatible with normal function and must not be interpreted as malfunctions.