Premium
Electrochemical Lateral Flow Devices: Towards Rapid Immunomagnetic Assays
Author(s) -
RuizVega Gisela,
Kitsara Maria,
Pellitero Miguel Aller,
Baldrich Eva,
del Campo F. Javier
Publication year - 2017
Publication title -
chemelectrochem
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.182
H-Index - 59
ISSN - 2196-0216
DOI - 10.1002/celc.201600902
Subject(s) - detection limit , chromatography , point of care , electrode , materials science , biomedical engineering , chemistry , medicine , nursing
Paper has been used for a long time in the production of low‐cost point‐of‐care (POC) detection devices, most of which rely on the detection of colored bands by the user. However, test strips rarely provide the quantitative detection of biomarkers in clinical samples. In contrast, electrochemical detection enables quantitative detection in paper‐based immunoassays. Here, we demonstrate the production of a simple and economical lateral‐flow electrode device using only screen printing. Contrary to previous work, this device is used to simplify the washing and detection of a short and sensitive immunomagnetic assay for myeloperoxidase (MPO) detection. As we show, after incubation of the sample with magnetic beads (MBs) and detection antibody (Ab) for only 5 min, the whole reaction mixture can be directly pipetted onto the strip. Washing and detection can then be performed, as the MBs are retained with the aid of a magnet, and minimal user intervention is needed. The obtained results reveal that MPO can be detected in 1 : 100 diluted serum with limits of detection and quantification of 0.18 and 0.62 ng mL −1 , respectively, in less than 15 min.