Recapture failure in transcatheter aortic valve replacement with CoreValve Evolut R
Author(s) -
Ito Nobuyasu,
Zen Kan,
Kuwabara Kensuke,
Matoba Satoaki
Publication year - 2021
Publication title -
catheterization and cardiovascular interventions
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.988
H-Index - 116
eISSN - 1522-726X
pISSN - 1522-1946
DOI - 10.1002/ccd.29580
Subject(s) - medicine , prosthesis , cardiology , stenosis , ascending aorta , surgery , aorta
Self‐expanding prostheses for transcatheter aortic valve replacement (TAVR), which can be recaptured, provide us the option of repositioning for a more accurate placement. We report a very rare case in which the recapture of CoreValve Evolut R (Medtronic, Minneapolis, Minnesota) to correct the implantation depth during the deployment could not be achieved. We planned TAVR with a 23 mm Evolut R prosthesis for a 92‐year‐old female with severe aortic stenosis and tightly bent thoracic aorta. During the first deployment attempt, the implantation depth was greater than we expected at 2/3 deployment. They tried to recapture and reposition the prosthesis, but the prosthesis was not re‐sheathed into the capsule of the delivery system. The prosthesis could not be recaptured despite a repeat attempt, and they were forced to deploy the device as it was. The prosthesis was deployed very carefully and implanted successfully without a pop‐up into the ascending aorta. At a later date, this situation was replicated in vitro and was found that the distal segment of the capsule became deformed, increasing the resistance to rotating the grip handle.