z-logo
Premium
Treatment of rotablation‐induced ostial left circumflex perforation by papyrus covered stent and its fenestration to recover the left anterior descending artery during CHIP procedure
Author(s) -
Adusumalli Srikanth,
Gaikwad Niranjan,
Raffel Chris,
Dautov Rustem
Publication year - 2019
Publication title -
catheterization and cardiovascular interventions
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.988
H-Index - 116
eISSN - 1522-726X
pISSN - 1522-1946
DOI - 10.1002/ccd.28114
Subject(s) - medicine , stent , percutaneous coronary intervention , lumen (anatomy) , circumflex , conventional pci , perforation , artery , cardiology , right coronary artery , surgery , cardiac tamponade , radiology , myocardial infarction , coronary angiography , metallurgy , punching , materials science
Coronary artery perforation is a rare complication of percutaneous coronary intervention (PCI). Covered stents have been successfully used in these situations. We report a case of ostial left circumflex (LCx) artery perforation during rotablation PCI of left main coronary artery (LMCA) and LCx artery. After failed attempts to balloon tamponade the perforation, a PK Papyrus covered stent was deployed from proximal LCx into LMCA. This resulted in acute exclusion of the left anterior descending (LAD) artery from coronary circulation. Using a dual lumen catheter, a stiff wire was advanced through the side port toward the occluded LAD to fenestrate the membrane of the covered stent. A series of balloons were used to dilate the fenestration in the covered stent to restore a normal flow into the LAD.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom