z-logo
Premium
Fluoroscopy‐guided snare retrieval of the celt ACD ® metallic vascular closure device following failed deployment
Author(s) -
Cahill Thomas J.,
Choji Kiyoshi,
Kardos Attila
Publication year - 2013
Publication title -
catheterization and cardiovascular interventions
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.988
H-Index - 116
eISSN - 1522-726X
pISSN - 1522-1946
DOI - 10.1002/ccd.25229
Subject(s) - medicine , vascular closure device , fluoroscopy , femoral artery , coronary angiography , closure (psychology) , angiography , radiology , vascular access , surgery , cardiology , market economy , hemodialysis , myocardial infarction , economics
We report a case of endovascular snare retrieval of a new stainless steel vascular closure device (Celt ACD ® , Kimal, Middlesex, UK) from the common femoral artery, following device failure after diagnostic coronary angiography. The stainless steel composition of the device aided successful fluoroscopic localization and removal. © 2013 Wiley Periodicals, Inc.