Abstract We report a case of dislodged and damaged stent during transradial coronary procedure using 6 Fr device, which was successfully retrieved by using a forcep and 8 Fr antegrade brachial sheath. The disfigured and bulky stent can be removed, after their retrieval from the coronary circulation, using a forcep inserted through an 8 Fr brachial artery sheath if the radial artery is deemed too small to accommodate larger sheath. Cathet Cardiovasc Intervent 2001;52:489–491. © 2001 Wiley‐Liss, Inc.