Ion‐exchange vs reversed‐phase chromatography for separation and determination of basic psychotropic drugs
Author(s) -
Petruczynik Anna,
Wróblewski Karol,
Deja Michał,
WaksmundzkaHajnos Monika
Publication year - 2015
Publication title -
biomedical chromatography
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.4
H-Index - 65
eISSN - 1099-0801
pISSN - 0269-3879
DOI - 10.1002/bmc.3482
Subject(s) - chemistry , chromatography , ion chromatography , ion exchange , ion , organic chemistry
Ion exchange chromatography, an alternative to reversed‐phase (RP) chromatography, is described in this paper. We aimed to obtain optimal conditions for the separation of basic drugs because silica‐based RP stationary phases show silanol effect and make the analysis of basic analytes hardly possible. The retention, separation selectivity, symmetry of peaks and system efficiency were examined in different eluent systems containing different types of buffers at acidic pH and with the addition of organic modifiers: methanol and acetonitrile. The obtained results reveal a large influence of the salt cation used for buffer preparation and the type of organic modifier on the retention behavior of the analytes. These results were also compared with those obtained on an XBridge C 18 column. The obtained results demonstrated that SCX stationary phases can be successfully used as alternatives to C 18 stationary phases in the separation of basic compounds. The most selective and efficient chromatographic systems were applied for the quantification of some psychotropic drugs in fortified human serum samples. Copyright © 2015 John Wiley & Sons, Ltd.