Premium
Enzyme activity deviates due to spatial and temporal temperature profiles in commercial microtiter plate readers
Author(s) -
Grosch JanHendrik,
Sieben Michaela,
Lattermann Clemens,
Kauffmann Kira,
Büchs Jochen,
Spieß Antje C.
Publication year - 2016
Publication title -
biotechnology journal
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.144
H-Index - 84
eISSN - 1860-7314
pISSN - 1860-6768
DOI - 10.1002/biot.201500422
Subject(s) - cuvette , microtiter plate , bioprocess , biological system , chemistry , absorbance , analytical chemistry (journal) , chromatography , chemical engineering , optics , physics , biology , engineering
Microtiter plates (MTP) and automatized techniques are increasingly applied in the field of biotechnology. However, the susceptibility of MTPs to edge effects such as thermal gradients can lead to high variation of measured enzyme activities. In an effort to enhance experimental reliability, to quantify, and to minimize instrument‐caused deviations in enzyme kinetics between two MTP‐readers, we comprehensively quantified temperature distribution in 96‐well MTPs. We demonstrated the robust application of the absorbance dye cresol red as easily applicable temperature indicator in cuvettes and MTPs and determined its accuracy to ±0.16°C. We then quantified temperature distributions in 96‐well MTPs revealing temperature deviations over single MTP of up to 2.2°C and different patterns in two commercial devices (BioTek Synergy 4 and Synergy Mx). The obtained liquid temperature was shown to be substantially controlled by evaporation. The temperature‐induced enzyme activity variation within MTPs amounted to about 20 %. Activity deviations between MTPs and to those in cuvettes were determined to 40 % due to deviations from the set temperature in MTPs. In conclusion, we propose a better control of experimental conditions in MTPs or alternative experimental systems for reliable determination of kinetic parameters for bioprocess development.