Subject(s) - derivative (finance) , phenylephrine , chromatography , hydrochloride , chemistry , medicine , business , organic chemistry , finance , blood pressure , radiology
Rapid, simple and sensitive derivative emission spectrofluorimetric methods have been developed for the simultaneous analysis of binary mixtures of guaifenesin (GUA) and phenylephrine hydrochloride (PHE). The methods are based upon measurement of the native fluorescence intensity of the two drugs at λ ex = 275 nm in methanolic solutions, followed by differentiation using first (D1) and second (D2) derivative techniques. The derivative fluorescence intensity–concentration plots were rectilinear over a range of 0.1–2 µg/mL for both GUA and PHE. The limits of detection were 0.027 (D1, GUA), 0.025 (D2, GUA), 0.031 (D1, PHE) and 0.033 (D2, PHE) µg/mL and limits of quantitation were 0.089 (D1, GUA), 0.083 (D2, GUA), 0.095 (D1, PHE) and 0.097 (D2, PHE) µg/mL. The proposed derivative emission spectrofluorimetric methods (D1 and D2) were successfully applied for the determination of the two compounds in binary mixtures and tablets with high precision and accuracy. The proposed methods were fully validated as per ICH guidelines. Copyright © 2014 John Wiley & Sons, Ltd.