SAR in rats exposed in 2,450‐MHz circularly polarized waveguides
Author(s) -
Chou ChungKwang,
Guy Arthur W.,
Johnson Robert B.
Publication year - 1984
Publication title -
bioelectromagnetics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.435
H-Index - 81
eISSN - 1521-186X
pISSN - 0197-8462
DOI - 10.1002/bem.2250050404
Subject(s) - specific absorption rate , absorption (acoustics) , physics , optics , waveguide , nuclear magnetic resonance , telecommunications , computer science , antenna (radio)
Average specific absorption rates (SARs) for live rats exposed in 2,450‐MHz circularly polarized waveguides were estimated from the total system loss determined from measurements using five power meters, and a correction factor representing actual SAR/apparent SAR. The actual SAR was measured by twin‐well calorimetry and the apparent SAR by power meters. Values were obtained for carcasses of various body masses for five orientations. The average SAR with free movement in the cages changed less than threefold as the rats grew from 200 to 700 g. The ratio of peak to average SAR in the body was less than 3. These results indicate relatively constant energy disposition in rats exposed in the circularly polarized waveguide.