z-logo
open-access-imgOpen Access
Variation in synovial fluid analysis by hospital laboratories
Author(s) -
Hasselbacher Peter
Publication year - 1987
Publication title -
arthritis & rheumatism
Language(s) - English
Resource type - Journals
eISSN - 1529-0131
pISSN - 0004-3591
DOI - 10.1002/art.1780300606
Subject(s) - synovial fluid , medicine , white blood cell , sodium , pathology , chemistry , osteoarthritis , alternative medicine , organic chemistry
A survey of 42 hospitals revealed that the frequency of synovial fluid analysis was low (mean 3.9 per month). The mean was skewed upward by a higher frequency in 3 teaching hospitals, such that the overall median was 1.5 per month, with 15 hospitals studying 1 or less per month. There did not appear to be any consensus regarding what constitutes a routine synovial fluid analysis. Technical errors occurred, including dilution with acetic acid for cell counts and inappropriate use of automated cell counters. Twenty‐six laboratories studied 4 “unknown” fluids for crystals. In 50 opportunities to detect each crystal type, calcium pyrophosphate was detected only 6 times, and sodium urate was detected 39 times. The white blood cell count reported for a single fluid ranged from 2, 467–12,000/mm 3 (mean ± SD 6,683 ± 1,992). The proportion of neutrophils identified varied from 0–30% (mean ± SD 13.8 ± 8.6). The majority of respondents believed that training and reference materials related to synovial fluid analysis should be improved.

The content you want is available to Zendy users.