z-logo
Premium
An evaluation of the DAWN‐B light scattering unit from wyatt technology: Suggested calibration, normalization, and clarification procedures
Author(s) -
Yunan Wan,
Zhongde Xu,
Li Jianbo,
Rosenblum William M.,
Mays Jimmy W.
Publication year - 1993
Publication title -
journal of applied polymer science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.575
H-Index - 166
eISSN - 1097-4628
pISSN - 0021-8995
DOI - 10.1002/app.1993.070490603
Subject(s) - photometer , normalization (sociology) , toluene , polystyrene , light scattering , calibration , tetrahydrofuran , scattering , analytical chemistry (journal) , optics , materials science , chemistry , chromatography , physics , organic chemistry , quantum mechanics , solvent , sociology , anthropology , polymer
Abstract Proper normalization, calibration, and sample clarification procedures for use with the Wyatt Technology DAWN‐B laser light scattering photometer are described. The utility of the proposed methods is confirmed by comparing results obtained on the solution properties of polystyrene standards in three different solvents (toluene, tetrahydrofuran, and methyl ethyl ketone) with results from the literature and generated using a Chromatix KMX‐6 low‐angle light scattering photometer. © 1993 John Wiley & Sons, Inc.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here