Interpretation of intrafraction motion review data and method for verification
Author(s) -
Graeper Gavin,
Cetnar Ashley,
Ayan Ahmet S.,
Weldon Michael
Publication year - 2021
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1002/acm2.13379
Subject(s) - interpretation (philosophy) , computer science , medical physics , medicine , programming language
The current clinical interface for Varian's intrafraction motion review (IMR) is limited, providing only qualitative data for review at the treatment console. This study provides a method of extracting and interpreting data from combined log files for quantitative evaluation. Combined log files acquired during patient treatment and a parsing code was developed to scan the combined log file looking for unique identifiers pertaining to the data of interest. We were able to extract clinically relevant parameters from the log files including date and time, gantry angle, expected marker position, found marker position, pixel size, and detection result. This study details how to compare IMR data to Calypso investigating dual‐surrogates for intrafraction monitoring during treatment for other researchers to build on these methods. Understanding data recorded during treatment within the combined log files can be helpful in quality improvement of patient care by retrospectively reviewing intrafraction motion.