Radiation treatment planning study to investigate feasibility of delivering I mmunotherapy in C ombination with A blative R adiosurgery to U ltra‐High Do S es (ICARUS)
Author(s) -
Rokni Michelle B.,
Pointer Kelli B.,
George Jonathan,
Luke Jason J.,
Chmura Steven J.,
Redler Gage
Publication year - 2021
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1002/acm2.13204
Subject(s) - medicine , nuclear medicine , ablative case , radiation therapy , radiosurgery , radiation treatment planning , tomotherapy , pembrolizumab , medical physics , radiology , cancer , immunotherapy
Purpose Immune checkpoint inhibitors improve survival in metastatic diseases for some cancers. Multisite SBRT with pembrolizumab (SBRT + Pembro) was shown to be safe with promising local control using biologically effective doses (BEDs) = 95–120 Gy. Increased BED may improve response rate; however, SBRT doses are limited by surrounding organs at risk (OARs). The purpose of this work was to develop and validate methods for safe delivery of ultra‐high doses of radiation (BED 10 > 300) to be used in future clinical trials. Methods and Materials The radiation plans from 15 patients enrolled on a phase I trial of SBRT + pembro were reanalyzed. Metastatic disease sites included liver (8/15), inguinal region (1/15), pelvis (2/15), lung (1/15), abdomen (1/15), spleen (1/15), and groin (1/15). Gross tumor volumes (GTVs) ranged from 80 to 708 cc. Following the same methodology used in the Phase I trial on which these patients were treated, GTVs > 65 cc were contracted to a 65 cc subvolume (SubGTV) resulting in only a portion of the GTV receiving prescription dose. Volumetric modulated arc therapy (VMAT) was used to plan treatments BED 10 = 360 Gy. Plans utilizing both 6FFF and 10FFF beams were compared to clinical plans delivering BED 10 = 112.50 Gy. The target primary goal was V100% > 95% with a secondary goal of V70% > 99% and OAR objectives per the trial. To demonstrate feasibility, plans were delivered to a diode array phantom and evaluated for fidelity using gamma analysis. Results All 30 plans met the secondary coverage goal and satisfied all OAR constraints. The primary goal was achieved in 12/15 of the 6FFF plans and 13/15 of the 10FFF plans. Average gamma analysis passing rate using criteria of 3% dose difference and 3, 2, and 1 mm were 99.1 ± 1.0%, 98.5 ± 1.6%, and 95.1 ± 3.8%, respectively. Conclusion Novel VMAT planning approaches with clinical treatment planning software and linear accelerators prove capable of delivering radiation doses in excess of 360 Gy BED 10 to tumor subvolumes, while maintaining safe OAR doses.