Experimental evaluation of the GE NM/CT 870 CZT clinical SPECT system equipped with WEHR and MEHRS collimator
Author(s) -
Ito Toshimune,
Matsusaka Yohji,
Onoguchi Masahisa,
Ichikawa Hajime,
Okuda Koichi,
Shibutani Takayuki,
Shishido Masaaki,
Sato Kozo
Publication year - 2021
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1002/acm2.13158
Subject(s) - collimator , nuclear medicine , imaging phantom , image resolution , scanner , spect imaging , physics , optics , materials science , medicine
Purpose A high‐energy‐resolution whole‐body SPECT‐CT device (NM/CT 870 CZT; C‐SPECT) equipped with a CZT detector has been developed and is being used clinically. A MEHRS collimator has also been developed recently, with an expected improvement in imaging accuracy using medium‐energy radionuclides. The objective of this study was to compare and analyze the accuracies of the following devices: a WEHR collimator and the MEHRS collimator installed on a C‐SPECT, and a NaI scintillation detector‐equipped Anger‐type SPECT (A‐SPECT) scanner, with a LEHR and LMEGP. Methods A line phantom was used to measure the energy resolutions including collimator characteristics in the planar acquisition of each device using 99m Tc and 123 I. We also measured the system's sensitivity and high‐contrast resolution using a lead bar phantom. We evaluated SPECT spatial resolution, high‐contrast resolution, radioactivity concentration linearity, and homogeneity, using a basic performance evaluation phantom. In addition, the effect of scatter correction was evaluated by varying the sub window (SW) employed for scattering correction. Results The energy resolution with 99m Tc was 5.6% in C‐SPECT with WEHR and 9.9% in A‐SPECT with LEHR. Using 123I, the results were 9.1% in C‐SPECT with WEHR, 5.5% in C‐SPECT with MEHRS, and 10.4% in A‐SPECT with LMEGP. The planar spatial resolution was similar under all conditions, but C‐SPECT performed better in SPECT acquisition. High‐contrast resolution was improved in C‐SPECT under planar condition and SPECT. The sensitivity and homogeneity were improved by setting the SW for scattering correction to 3% of the main peak in C‐SPECT. Conclusion C‐SPECT demonstrates excellent energy resolution and improved high‐contrast resolution for each radionuclide. In addition, when using 123I, careful attention should be paid to SW for scatter correction. By setting the appropriate SW, C‐SPECT with MEHRS has an excellent scattered ray removal effect, and highly homogenous imaging is possible while maintaining the high‐contrast resolution.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here