Initial testing of pegfilgrastim (Neulasta Onpro) on‐body injector in multiple radiological imaging environments
Author(s) -
Long Zaiyang,
Kurup Anil Nicholas,
Jensen Nicole M.,
Hangiandreou Nicholas J.,
Schueler Beth A.,
Yu Lifeng,
Leng Shuai,
Wood Christopher P.,
Felmlee Joel P.
Publication year - 2021
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1002/acm2.13156
Subject(s) - imaging phantom , scanner , ultrasound , medicine , magnetic resonance imaging , nuclear medicine , radiological weapon , radiology , biomedical engineering , medical physics , physics , optics
Abstract Purpose An increasing number of implantable or external devices can impact whether patients can receive radiological imaging examinations. This study examines and tests the Neulasta (pegfilgrastim) Onpro on‐body injector in multiple imaging environments. Methods The injector was analyzed for four imaging modalities with testing protocols and strategies developed for each modality. In x‐ray and computed tomography (CT), scans with much higher exposure than clinical protocols were performed with the device attached to an anthropomorphic phantom. The device was monitored until the completion of drug delivery. For magnetic resonance imaging (MRI), the device was assessed using a hand‐held magnet and underwent the magnetically induced displacement testing in a 1.5T clinical MRI scanner room. For ultrasound, magnetic field changes were measured around an ultrasound scanner system with three transducers. Results For x‐ray and CT no sign of device error was identified during or after the high radiation exposure scans. Drug delivery was completed at expected timing with expected volume. For MRI the device showed significant attractive force towards the hand‐held magnet and a 50‐degree deflection angle at 50 cm from the opening of the scanner bore. No further assessment from the gradient or radiofrequency field was deemed necessary. For ultrasound the maximum magnetic field change from baseline was measured to be +11.7 μT in comparison to +74.2 μT at 4 inches from a working microwave. Conclusions No device performance issue was identified under the extreme test conditions in x‐ray or CT. The device was found to be MR Unsafe. Magnetic field changes around an ultrasound system met the limitation set by manufacture. Patient ultrasound scanning is considered safe as long as the transducers do not inadvertently loosen the device.