Recommendations of megavoltage computed tomography settings for the implementation of adaptive radiotherapy on helical tomotherapy units
Author(s) -
Velten Christian,
Boyd Robert,
Jeong Kyoungkeun,
Garg Madhur K.,
Tomé Wolfgang A.
Publication year - 2020
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1002/acm2.12859
Subject(s) - tomotherapy , nuclear medicine , image quality , imaging phantom , iterative reconstruction , contrast to noise ratio , image resolution , tomography , medicine , radiation therapy , materials science , optics , physics , radiology , computer science , artificial intelligence , image (mathematics)
Megavoltage computed tomography (MVCT) image quality metrics were evaluated on an Accuray Radixact unit to recommend scan settings for the implementation of a consistent adaptive radiotherapy program. Megavoltage computed tomography image quality was evaluated and compared to a kilovoltage CT (kVCT) simulator using a commercial cone beam computed tomography image quality phantom. Megavoltage computed tomographies were acquired on the Accuray Radixact using fine, normal, and coarse pitches, with all available reconstruction slice thicknesses, each of which were reconstructed using standard and iterative reconstruction (IR). Image quality metrics (IQM) were evaluated using DoseLab: automatically and manually calculated spatial resolution, subject contrast, and contrast‐to‐noise ratio (CNR). Scanning time was 15.6 s/cm for fine, 8.1 s/cm for normal, and 5.6 s/cm for coarse pitch. Automatically evaluated spatial resolutions ranged from 0.39, 0.41, to 0.42 lp/mm for standard reconstruction and from 0.24, 0.21, to 0.18 lp/mm for soft‐tissue IR, respectively, with general IR yielding values in between these. Spatial resolution for kVCT was measured to be at least 0.42 lp/mm. Contrast was consistent across MVCT settings with 8.1 ± 0.2%, while kVCT contrast was 10.27 ± 0.05%. CNR was calculated to be 3.3 ± 0.4 for standard reconstruction, 7.4 ± 0.4 for general IR, and 12.0 ± 1.9 for soft‐tissue IR. It was found that increasing reconstruction slice thickness for a given pitch does not improve IQMs. Based on the consistency of contrast metrics across pitch values and the only slightly reduced spatial resolution using normal compared to fine pitch, we recommend the use of normal pitch with 2 mm slice thickness to maximize image quality for ART while limiting scanning time. Only for sites for which improved CNR is required and reduced spatial resolution is acceptable, soft‐tissue IR is recommended.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here