Intrafractional relationship changes between an external breathing signal and fiducial marker positions in pancreatic cancer patients
Author(s) -
Pettersson Niclas,
Oderinde Oluwaseyi M.,
Murphy James,
Simpson Daniel,
Cerviño Laura I.
Publication year - 2020
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1002/acm2.12841
Subject(s) - isocenter , fiducial marker , medicine , nuclear medicine , breathing , radiation therapy , pancreatic cancer , radiology , cancer , anatomy
Background and purpose The purpose of this study of pancreatic cancer patients treated with respiratory‐guided stereotactic body radiotherapy (SBRT) on a standard linac was to investigate (a) the intrafractional relationship change (IRC) between a breathing signal and the tumor position, (b) the impact of IRC on the delivered dose, and (c) potential IRC predictors. Materials and methods We retrospectively investigated 10 pancreatic cancer patients with 2–4 implanted fiducial markers in the tumor treated with SBRT. Fluoroscopic images were acquired before and after treatment delivery simultaneously with the abdominal breathing motion. We quantified the IRC as the change in fiducial location for a given breathing amplitude in the left–right (LR), anterior–posterior (AP), and superior–inferior (SI) directions from before to after treatment delivery. The treatment plans were re‐calculated after changing the isocenter coordinates according to the IRCs. Four treatment‐ or patient‐related factors were investigated as potential predictors for IRC using linear models. Results The average (±1 SD) absolute IRCs in the LR, AP, and SI directions were 1.2 ± 1.2 mm, 0.7 ± 0.7 mm, and 1.1 ± 0.8 mm, respectively. The average 3D IRC was 2.0 ± 1.3 mm (range: 0.4–5.3 mm) for a median treatment delivery time of 8.5 min (range: 5.7–19.9 min; n = 31 fractions). The dose coverage of the internal target volume (ITV) decreased by more than 3% points in three of 31 fractions. In those cases, the 3D IRC had been larger than 4.3 mm. The 3D IRC was found to correlate with changes in the minimum breathing amplitude during treatment delivery. Conclusion On average, 2 mm of treatment delivery accuracy was lost due to IRC. Periodical intrafractional imaging is needed to safely deliver respiratory‐guided SBRT.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here