Subatmospheric electrospray interface for coupling of microcolumn separations with mass spectrometry
Author(s) -
Foret Frantisek,
Zhou Haihong,
Gangl Eric,
Karger Barry L.
Publication year - 2000
Publication title -
electrophoresis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.666
H-Index - 158
eISSN - 1522-2683
pISSN - 0173-0835
DOI - 10.1002/(sici)1522-2683(20000401)21:7<1363::aid-elps1363>3.0.co;2-u
Subject(s) - electrospray , chromatography , chemistry , analytical chemistry (journal) , mass spectrometry , capillary action , volumetric flow rate , body orifice , capillary electrophoresis–mass spectrometry , electrospray ionization , materials science , mechanics , ecology , physics , composite material , biology
A modular subatmospheric electrospray interface with fiber optic UV detection close to the electrospray tip was developed for coupling of microcolumn separation techniques with mass spectrometry. The interface was based on a liquid junction with a removable microelectrospray tip. The electrospray tip was enclosed in a subatmospheric chamber attached in front of the sampling orifice of the mass spectrometer. The inlet of the liquid junction was maintained at atmospheric pressure, and thus no pressure drop developed across the separation column. The flow rate of the electrosprayed liquid from the liquid junction reservoir was adjusted by the pressure in the electrospray chamber. In this approach, a continuous and stable electrospray could be achieved without the use of an external pump. Since the electrospray did not depend on fluid delivery from the separation column, coated capillaries without electroosmotic flow as well as capillaries with electroosmotic flow could be used for capillary electrophoresis. In addition, the interface was found to be effective with capillary liquid chromatography. The use of a fiber optic UV detector placed close to the exit of the separation column provided additional detection information and a simple means of troubleshooting. The interface did not significantly influence the quality of the separation, even with columns generating several hundred thousand theoretical plates. Peptide samples in the submicromolar concentration range were detected, corresponding to a limit of detection in the attomole range.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here