Improved near-infrared spectral responsivity scale
Author(s) -
Ping-Shine Shaw,
Thomas C. Larason,
Rajeev Gupta,
Steven W. Brown,
Keith R. Lykke
Publication year - 2000
Publication title -
journal of research of the national institute of standards and technology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.202
H-Index - 59
eISSN - 2165-7254
pISSN - 1044-677X
DOI - 10.6028/jres.105.055
Subject(s) - monochromator , responsivity , radiometer , optics , photodiode , materials science , optoelectronics , photodetector , detector , wavelength , physics
A cryogenic radiometer-based system was constructed at the National Institute of Standards and Technology for absolute radiometric measurements to improve detector spectral power responsivity scales in the wavelength range from 900 nm to 1800 nm. In addition to the liquid-helium-cooled cryogenic radiometer, the system consists of a 100 W quartz-tungsten-halogen lamp light source and a 1 m single-grating monochromator for wavelength selection. The system was characterized and the uncertainty in spectral power responsivity measurements evaluated. A variety of photodetectors, including indium gallium arsenide photodiodes (InGaAs), germanium (Ge) photodiodes, and pyroelectric detectors, were subsequently calibrated. Over most of the spectral range, the spectral power responsivity of the photodetectors can be measured with a combined relative standard uncertainty of 0.4 % or less. This is more than a factor of two smaller than our previous capabilities, and represents a significant improvement in the near infrared (NIR) spectral power responsivity scale maintained at NIST. We discuss the characterization of the monochromator-based system and present results of photodetector spectral power responsivity calibrations.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom