Quality by Design based Developed and Validation of RP-HPLC Method for Simultaneous Estimation of Pazopanib in Bulk and Pharmaceutical Dosage Forms
Author(s) -
Kiran Kumar Buralla,
V. Parthasarathy
Publication year - 2019
Publication title -
international journal of pharmaceutical investigation
Language(s) - English
Resource type - Journals
eISSN - 2230-973X
pISSN - 2230-9713
DOI - 10.5530/ijpi.2019.3.25
Subject(s) - chromatography , quality by design , dosage form , relative standard deviation , computer science , phosphate buffered saline , gradient elution , high performance liquid chromatography , materials science , chemistry , detection limit , particle size
Objectives: This paper describes the development of an accurate, precise, robust, sensitive, economical and rapid isocratic RP-HPLC technique complying quality by design and validate according to ICH guidelines. The simultaneous estimation of the Pazopanib with Dasatinib as internal standard in bulk and pharmaceutical dosage forms with the help of chemo metrics, multicriteria decision-making approach. The main objective was to identify the robust chromatographic conditions where an adequate separation of the components with quality peaks, within acceptable run time can be achieved. Target Analytical Profile (TAP) was defined and systematic risk analysis was carried out to identify Critical Method Attributes (CMA) having an impact on Critical Quality Attributes (CQA). % Acetonitrile (ACN), KH2PO4, pH and Flow rate was identified as CMA. Methods: The effective chromatographic separation was achieved by utilizing Phenomenex Enable C18 column and mobile phase consisted of ACN and Phosphate buffer (60:40 %v/v, pH 5) adjusted with orthophosphoric acid and the flow rate of 1.2 ml/min. the elution was monitored at 290nm using a PDA-UV detector. Results: The developed method resulted in eluting the drug at 2.190 min, respectively. The regression coefficients were observed to be 0.998 for all models. The LOD was about 10.74 nano-gram/ml and LOQ were about 31.74 nano-gram/ml. The relative standard deviation was observed to be 0.5079%. Conclusion: The results of the study demonstrated that the suggested RP-HPLC method is rapid, simple, accurate and precise. The method was validated by determining its accuracy, precision and system stability.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom