Inappropriate Asystole Detection in Early Postoperative Phase after Loop Recorder Implantation
Author(s) -
Miriam Bortnik,
Eraldo Occhetta,
Andrea Magnani,
Anna Degiovanni,
Paolo Marino
Publication year - 2011
Publication title -
isrn cardiology
Language(s) - English
Resource type - Journals
eISSN - 2090-5599
pISSN - 2090-5580
DOI - 10.5402/2011/146062
Subject(s) - asystole , implantable loop recorder , medicine , cardiology , anesthesia , atrial fibrillation
The implantable loop recorder is a useful diagnostic tool for patients with unexplained syncope. The capability to automatically detect and store arrhythmic events, implemented in the last generations of these devices, can further improve the diagnostic yield, but this feature can be compromised by inappropriate detection of false arrhythmias. We herein report the case of a patient in which several inappropriate activations of long-lasting asystole occurred in the two days following the implant, probably because of an intermittently loose contact between the device and subcutaneous tissue for a small pocket haematoma.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom