Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
Author(s) -
Antoine Kossaify,
Nayla Nicolas
Publication year - 2012
Publication title -
clinical medicine insights case reports
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.187
H-Index - 12
ISSN - 1179-5476
DOI - 10.4137/ccrep.s9625
Subject(s) - ventricle , lead (geology) , cardiology , medicine , backup , ventricular pacing , heart failure , computer science , geology , operating system , geomorphology
An 80-year-old patient having a dual chamber DR pacer connected to a VDD lead presented with chronic lead dislodgment with the atrial ring displaced into the right ventricle. There was no ventricular capture at maximal ventricular output, and given the clinical settings, the condition was managed with a conservative approach, the mode was switched to AAI ensuring a minimal adequate ventricular pacing backup in a non dependent patient.