Blood Pressure Measurement in Peritoneal Dialysis: Which Method is Best?
Author(s) -
O'Shaughnessy Michelle M.,
Durcan Martin,
Kinsella Sinead M.,
Griffin Matthew D.,
Reddan Donal N.,
Lappin David W.
Publication year - 2013
Publication title -
peritoneal dialysis international
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.79
H-Index - 83
eISSN - 1718-4304
pISSN - 0896-8608
DOI - 10.3747/pdi.2012.00027
Subject(s) - medicine , ambulatory blood pressure , blood pressure , ambulatory , confidence interval , population , masked hypertension , white coat hypertension , bland–altman plot , cardiology , limits of agreement , surgery , nuclear medicine , environmental health
Background The optimal approach to monitoring blood pressure (BP) in the peritoneal dialysis (PD) population is unclear. Ambulatory BP monitoring reliably predicts prognosis, but can be inconvenient. The accuracy of home BP monitoring in this population is unproven. The automated BpTRU device (BpTRU Medical Devices, Coquitlam, BC, Canada), which provides an average of up to 6 successive in-office BP measurements, has not been studied in this patient group.Methods We studied 17 patients (average age: 54 ± 12 years; 12 men, 5 women; 94% on automated PD) attending a single center. All patients underwent office, home, BpTRU, and ambulatory BP measurement. The reference standard for analysis was daytime ambulatory BP. Correlation between the referent method and each comparator method was determined (Pearson correlation coefficient), and Bland–Altman scatter plots depicting the differences in the BP measurements were constructed.Results Mean office BP (126.4 ± 16.9/78.8 ± 11.6 mmHg) and BpTRU BP (123.8 ± 13.7/80.7 ± 11.1 mmHg) closely approximated mean daytime ambulatory BP (129.3 ± 14.8/78.2 ± 7.9 mmHg). Mean home BP (143.8 ± 15.0/89.9 ± 28.1 mmHg) significantly overestimated mean daytime systolic BP by 14.2 mmHg (95% confidence interval: 4.3 mmHg to 24.1 mmHg; p = 0.008). Bland–Altman plots demonstrated poorest agreement between home BP and daytime ambulatory BP. No patient had “white-coat hypertension,” and only 1 patient had false-resistant hypertension. Most patients showed abnormal nocturnal dipping patterns (non-dipping: n = 11; reverse-dipping: n = 5; normal dipping: n = 1).Conclusions We report a novel finding that BP measurement using the BpTRU device is more accurate than home BP measurement in a PD population. Potential explanations for this observation include poor home BP measurement technique, use of poorly validated home BP measurement devices, or a reduced prevalence of white-coat effect among PD patients. Our study also confirms that, in the PD population, BP measurements vary considerably with patient location, time of day, and measurement technique.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom