Koszorúérsztentek fémmel fedett felületének vizsgálata
Author(s) -
Dóra Károly,
Miksa Kovács,
Attila Terdik,
Eszter Bognár,
János Dobránszky
Publication year - 2013
Publication title -
fiatal műszakiak tudományos ülésszaka
Language(s) - English
Resource type - Journals
eISSN - 2668-3792
pISSN - 2067-6808
DOI - 10.36243/fmtu-2013.35
Subject(s) - physics , mathematics
During our work we compared two examination methods used for investigating the metallic surface area (MSA) of coronary stents. Each measurement process was tried on different type of stents. Both methods aimed at converting the cylindrical stent into a flattened two-dimensional image. The first method was manually performed using a stereomicroscope. The second method, which has recently been developed, is an automatized method using an integrated scanner and rotating engine. We also determined the largest and the smallest cell sizes from these flattened stent pattern. Our findings showed the largest difference between the two methods was the speed and the accuracy.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom