z-logo
open-access-imgOpen Access
Increasing the Low-Glucose Alarm of a Continuous Glucose Monitoring System Prevents Exercise-Induced Hypoglycemia Without Triggering Any False Alarms
Author(s) -
Katherine E. Iscoe,
Raymond J. Davey,
Paul Fournier
Publication year - 2011
Publication title -
diabetes care
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 6.636
H-Index - 363
eISSN - 1935-5548
pISSN - 0149-5992
DOI - 10.2337/dc10-2243
Subject(s) - hypoglycemia , continuous glucose monitoring , alarm , medicine , diabetes mellitus , blood glucose self monitoring , incidence (geometry) , false alarm , type 1 diabetes , endocrinology , computer science , artificial intelligence , mathematics , materials science , composite material , geometry
The use of continuous glucose monitoring systems (CGMSs) with low-glucose alarms is advocated as a means to decrease the risk of hypoglycemia in type 1 diabetes. Unfortunately, marked mismatches between CGMS readings and actual blood glucose (BG) concentrations limit the usefulness of CGMS in preventing hypoglycemia (1). Although we showed recently that raising the alarm level to compensate for this mismatch decreases the incidence and duration of hypoglycemic episodes, this results in an unacceptably high rate of false alarms (1), defined as an alarm triggered when BG levels are greater than the alarm threshold. This is an important issue because repeated exposure to false alarms can discourage individuals from using their CGMSs (2). Given that CGMSs overestimate BG levels when they rapidly decline (3,4), we propose that raising the CGMS alarm …

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom