Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Author(s) -
Boris Kovatchev,
Stacey M. Anderson,
Lutz Heinemann,
William L. Clarke
Publication year - 2008
Publication title -
diabetes care
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 6.636
H-Index - 363
eISSN - 1935-5548
pISSN - 0149-5992
DOI - 10.2337/dc08-1303
Subject(s) - medicine , continuous glucose monitoring , diabetes mellitus , endocrinology , type 1 diabetes
In their comments on our assessment of the accuracy of four continuous glucose monitoring (CGM) sensors, Brauker and Matsubara (1) correctly point out that our presented studies employed hyperinsulinemic-euglycemic and hypoglycemic clamps, which tested the four devices (Guardian, DexCom 3-day sensor, Navigator, and Glucoday) close to the limits of their stated glucose range. We agree that such a test goes beyond evaluation of the devices per their …