Low-dose digital computed radiography in pediatric chest imaging
Author(s) -
Kogutt,
J.A. Jones,
D D S Perkins
Publication year - 1988
Publication title -
american journal of roentgenology
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 1.294
H-Index - 196
eISSN - 1546-3141
pISSN - 0361-803X
DOI - 10.2214/ajr.151.4.775
Subject(s) - medicine , radiography , computed radiography , digital radiography , radiology , nuclear medicine , computed tomography , medical physics , medical imaging , image quality , computer vision , computer science , image (mathematics)
A prototype digital computed radiographic imaging system that uses laser-stimulated luminescence was evaluated for its ability to achieve reproducible, high-detail, low-dose pediatric chest radiographs. Using this system, we performed a total of 401 examinations in infants and children, and achieved an 85% reduction in radiation dose, as compared with that delivered when film-screen techniques were used. We also achieved satisfactory image resolution, and the images obtained were of acceptable diagnostic quality. A direct comparison of analog and digital radiographs showed that comparable quality and clinical acceptability could be readily maintained between the two techniques. This study shows that high-quality images can be produced by this system at radiation doses reduced by 85% when compared with doses from standard radiographic techniques.