Validating a Stability Indicating HPLC Method for Kinetic Study of Cetirizine Degradation in Acidic and Oxidative Conditions
Author(s) -
Effat Souri,
Ali Hatami,
Nazanin Shabani Ravari,
Farhad Alvandifar,
Maliheh Barazandeh Tehrani
Publication year - 2013
Publication title -
iranian journal of pharmaceutical research : ijpr
Language(s) - English
DOI - 10.22037/ijpr.2013.1312
A stability indicating High-Performance Liquid Chromatography (HPLC) method was validated and used to study the degradation of cetirizine dihydrochloride in acidic and oxidative conditions. The separation was carried out on a Symmetry C18 column and a mixture of 50 mM KH2PO4 and acetonitrile (60:40 v/v, pH = 3.5) was used as the mobile phase. The method was linear over the range of 1-20 μg/mL of cetirizine dihydrochloride (r2 > 0.999) and the within-day and between-day precision values were less than 1.5%. The results showed that cetirizine dihydrochloride was unstable in 2 M HCl and 0.5% H2O2. The kinetics of the acidic degradation showed a pseudo-first-order reaction in the temperature range of 70-90°C. In addition, the kinetics of hydrogen peroxide mediated degradation was pseudo-first-order in the temperature range of 50-80°C.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom