Rapid High Performance Liquid Chromatographic Method for Determination of Clarithromycin in Human Plasma Using Amperometric Detection: Application in Pharmacokinetic and Bioequivalence Studies
Author(s) -
Seyed Mohsen Foroutan,
Afshin Zarghi,
Alireza Shafaati,
Babak Madadian,
Farshid Abolfathi
Publication year - 2013
Publication title -
iranian journal of pharmaceutical research : ijpr
Language(s) - English
DOI - 10.22037/ijpr.2013.1273
A rapid, sensitive and reproducible HPLC method using amperometric detector was developed and validated for the analysis of clarithromycin in human plasma. The separation was achieved on a monolithic silica column (MZ- C8 125×4.0 mm) using acetonitrile-methanol-potassium dihydrogen phosphate buffer (40:6:54,v/v), with pH of 7.5, as the mobile phase at a flow rate of 1.5 mL/min. The assay enables the measurement of clarithromycin for therapeutic drug monitoring with a minimum quantification limit of 20 ng/mL. The method involves simple, protein precipitation procedure and analytical recovery was complete. The calibration curve was linear over the concentration range of 0.1-6 μg/mL. The coefficients of variation for inter-day and intra-day assay were found to be less than 6%. This method was used in bioequivalency and pharmacokinetic studies of the test (generic) product 2 × 500 mg clarithromycin tablets, with respect to the reference product.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom