z-logo
open-access-imgOpen Access
Performance acceptance test of a portable instrument to detect uranium in water at the DOE Advanced Waste Water Treatment Plant, Fernald, Ohio
Author(s) -
Laura N. Anderson,
Stephan Weeks
Publication year - 1997
Language(s) - English
Resource type - Reports
DOI - 10.2172/537304
Subject(s) - environmental science , sample (material) , uranium , detection limit , sampling (signal processing) , chromatography , materials science , chemistry , computer science , metallurgy , filter (signal processing) , computer vision
The Eppendorf-Biotronik Model IC 2001-2, a portable field ruggedized ion chromatography instrument, was rigorously tested at the DOE Advanced Waste Water Treatment Plant, Fernald, Ohio. This instrument rapidly detected the uranium concentration in water, and has a detection limit in the low ppb range without using the sample concentrating feature. The test set of samples analyzed included: ``Real World`` water samples from the AWWT containing uranium concentrations in the 9--110 ppb range, a sample blank, and a performance evaluation sample. The AWWT samples contained sets of both raw water and acid-preserved water samples. Selected samples were analyzed in quadruplicate to asses the instrument`s precision, and these results were compared with the results from an off-site confirmatory laboratory to assess the instrument`s accuracy. Additional comparisons with on-site laboratory instruments, Chemcheck KPA-11 and Scintrex UA-3 are reported. Overall, the Eppendorf-Biotronik IC 2001-2 performed exceptionally well providing a detection limit in the low ppb region (< 10 ppb) and giving rapid (< 5 minutes) accurate and reproducible analytical results for the AWWT, ``real world``, water samples with uranium concentrations in the region of interest (10--40 ppb). The per sample operating cost for this instrument is equivalent to the per sample cost for the currently used KPA. The time required to analyze a sample and provide a result is approximately the same for the CI 2001-2, KPA, and Scintrex instruments

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here