Continued studies of calorimeter performance at the Lawrence Livermore Laboratory
Author(s) -
S.A. Steward,
R.T. Tsugawa
Publication year - 1975
Publication title -
osti oai (u.s. department of energy office of scientific and technical information)
Language(s) - English
Resource type - Reports
DOI - 10.2172/4134629
Subject(s) - calorimeter (particle physics) , thermopile , nuclear engineering , nuclear physics , physics , environmental science , radiochemistry , chemistry , detector , optics , engineering , infrared
Calibrations of two calorimeters used for tritium and plutonium assays were made. Data from three new standards of about 0.5, 1, and 5 W were added to the results of a previous report and analyzed together. The accuracies of both calorimeters appear to fall within the specified 0.5 percent, although the data now available for the large calorimeter is insufficient to permit a more definite conclusion. An expression of the bias correction for each calorimeter with respect to the sample power cannot be determined. The bias of the medium thermopile-type calorimeter tends to be positive, however, and that of the large resistance-bridge design appears to be negative. (auth
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom