BetaScint{trademark} fiber-optic sensor for detecting strontium-90 and uranium-238 in soil. Innovative technology summary report
Publication year - 1998
Publication title -
osti oai (u.s. department of energy office of scientific and technical information)
Language(s) - English
Resource type - Reports
DOI - 10.2172/354883
Subject(s) - collimated light , soil survey , sampling (signal processing) , detector , stack (abstract data type) , environmental science , remote sensing , physics , soil water , computer science , geography , optics , soil science , operating system , laser
Accurate measurements of radioactivity in soils contaminated with Strontium-90 (Sr-90) or Uranium-238 (U-238) are essential for many DOE site remediation programs. These crucial measurements determine if excavation and soil removal is necessary, where remediation efforts should be focused, and/or if a site has reached closure. Measuring soil contamination by standard EPA laboratory methods typically takes a week (accelerated analytical test turnaround) or a month (standard analytical test turnaround). The time delay extends to operations involving heavy excavation equipment and associated personnel which are the main costs of remediation. This report describes an application of the BetaScint{trademark} fiber-optic sensor that measures Sr-90 or U-238 contamination in soil samples on site in about 20 minutes, at a much lower cost than time-consuming laboratory methods, to greatly facilitate remediation. This report describes the technology, its performance, its uses, cost, regulatory and policy issues, and lessons learned

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom